R&S AFQ100A
General Instrument Settings
1401.3084.32
4.83
E-3
Setup Menu
This section describes the settings which do not directly affect the signal generation. They affect various
functions, such as storing instrument settings or setting the IEC/IEEE bus address of the instrument.
Most of these settings can only be accessed by means of dialogs which are opened using a keyboard.
Selecting
Setup
in the windows toolbar opens the
Setup
menu used to set the general instrument
parameters and to configure the instrument interfaces.
Internal Adjustments- Setup System
For calibration the
Internal Adjustments
Dialog can be selected in the
System
section of the
Setup
menu.
The R&S AFQ I/Q Modulation Generator is extremely accurate thanks to the integrated procedures for
adjustments.
All internal adjustments for which no external measuring equipment is needed can be started in the
Internal Adjustments
dialog. The adjustments with external measuring equipment are described in the
s
ervice
m
anual.
Adjustment is recommended if the temperature range in which the instrument is operated changes, or
prior to all applications which require maximum accuracy.
During adjustment a bar indicates the status of progress. The result is indicated in the dialog (Pass or
Failed). If an error occurs, adjustment is terminated and an error message is output in the info line.
Note:
Make sure that the instrument has reached the operating temperature before performing
adjustments. The warm-up time amounts 30 minutes.