![background image](http://html1.mh-extra.com/html/rands/afq100a/afq100a_operating-manual_3076923212.webp)
R&S AFQ100A
General Instrument Settings
1401.3084.32 4.101
E-3
Note:
Since
Factory Preset
resets the remote channel and network settings to the default values,
executing Factory Preset via remote control terminates the connection to the instrument, if these
settings had been configured to values different to the default ones!
The
Factory Preset
function resets nearly all instrument settings. In addition to the regular preset by
means of the
PRESET
key, a
Factory Preset
resets also the following values:
Reference frequency settings (Ref Oscillator dialog)
Network settings including hostname (Setup dialog)
Remote Channel settings including IEC/IEEE bus address and emulation (Setup dialog)
Start/Stop Gui Update (Setup dialog)
Display and keyboard settings (Setup dialog).
To maintain security, password settings and all settings protected by these passwords like disabled USB
and LAN connections are not changed.
Not affected by the
Factory Preset
are also user data, lists or instrument settings files, created for
example by means of the Save/Recall function.
Remote-control command
:
SYST:FPR
Help Menu
The help system offers comprehensive online help for the instrument. A desired topic can be selected
via the table of contents (select
Manual
) or the index (select
Index
).
For context-sensitive information about a marked parameter, press the
F1
key.