Chapter 6 Parameter Listing and Descriptions
Application Parameters
Operation
P100 Mode of Operation
This parameter sets the mode of operation, when in run mode, and can be set
to one of the following:
Option
Description
1 = Distance (Default)
Display shows the distance from the
transducer face to the surface.
2 = Level
Display shows how full the vessel is.
3 = Space
Display shows how empty a vessel is.
4 = OCM Head
Display shows how high the head is.
5 = OCM Flow
Display shows the instantaneous flow.
P101 Xducer (Transducer)
This parameter should be set to the transducer being used with the unit, and
can be set to one of the following:
Option
Description
1 = dB3
(Default)
Transducer is a dB3. Range 4.921 to 118.116
inches (inclusive of Far End Blanking).
2 = dB6
Transducer is dB6 Range 11.811 to 236.22 inches
3= dB10
Transducer is dB10 Range 11.811 to 393.701 inches
feet
4= dB15
Transducer is dB15 Range 19.685 to 590.551 inches
7 = dBS6
Transducer is dBS6 Range 7.874 to 236.22 inches
8 = dBMach3
Transducer is dBMach3 Range 0.0 to 95.472 inches
P102 Material
This parameter should be set to the type of material being monitored.
Option
Description
1 = Liquid (Default)
Use for liquids and flat solid materials
2 = Solid
Solid material that is heaped or at an angle
3 = Closed Tank
Use for applications within a closed tank or
where a secondary echo response may
become focussed to create a larger echo
than the first.
Summary of Contents for FlowCERT LITE
Page 1: ...FlowCERT LITE UL INSTRUCTION MANUAL...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...Page 6 EC DECLARATION OF CONFORMITY...
Page 18: ...Page 10 The full dimensions of the enclosure are as shown below...
Page 37: ...Page 29 This page left blank intentionally...
Page 77: ...Page 69...
Page 135: ...Page 127 This page left blank intentionally...