Page
101
Failsafe
P808 Fail-safe Mode
By default, if a fail-safe condition occurs, then the display, relays and the
mA output are held at their last
known
values until a valid reading is
obtained.
If required, then you can change this so that the unit goes to
high
(100% of
span), or
low
(empty) as follows:
Option
Description
1 = Known (Default)
Remain at the last
known
value
2 = High
Will fail to the
high
value (100% of Span).
3= Low
Will fail to the
low
value (empty)
See Also P218, P228, P238, P248, P258 - Relay Fail-safe and P840 mA Output Fail-safe
Important Information
In the event of a
fail-safe
condition occurring, the display, relays and mA
Output can be configured to fail to a condition which is independent of each
other. To set independent
Relay Failsafe
see
P218, 228, 238, 248, 258
. And
for independent
mA Output Failsafe
see
P840.
P809 Fail-safe Time
In the event of a fail-safe condition the fail safe timer determines the time
before fail-safe mode is activated.
Default = 2mins
If the timer activates, the unit goes into
fail-safe
, as determined by
P808
(
Display
),
P218, 228, 238, 248, 258
(
Relays
) and
P840
(
mA Output
).
When this happens, you will see the message “
Failed Safe
!” on the display,
along with a message explaining why (lost echo or transducer fault, for
example)
When a valid measurement is obtained then the display, relays and mA
output will be restored and the timer is reset.
Summary of Contents for FlowCERT LITE
Page 1: ...FlowCERT LITE UL INSTRUCTION MANUAL...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...Page 6 EC DECLARATION OF CONFORMITY...
Page 18: ...Page 10 The full dimensions of the enclosure are as shown below...
Page 37: ...Page 29 This page left blank intentionally...
Page 77: ...Page 69...
Page 135: ...Page 127 This page left blank intentionally...