Page 16
Important Information
If the equipment is installed or used in a manner not specified in this
manual, then the protection provided by the equipment may be
impaired.
Preparation for Operation
Before switching on, check the following:
FlowCERT
LITE
is mounted correctly and is in a ‘safe’ area.
The power supply is correctly installed.
The relays are connected correctly.
Important Information
Before applying AC power (mains), make sure the supply is 115V
AC.
Never operate the FlowCERT L
ITE
with terminal access exposed.
An external switch or circuit breaker should be installed near to the
FlowCERT L
ITE
to allow the supply to be removed during
installation and maintenance. In addition, the relay contacts should
also have a means of isolating them from the FlowCERT L
ITE
.
Interconnecting cables must be adequately insulated in accordance
with local regulations. Strip back 30 mm of the outer insulation of
the cable. Strip 5 mm of insulation from the end of each conductor.
Twist all exposed strands of the conductor together. Insert the
stripped conductor into the terminal block as far as it will go and
tighten the terminal block screw. Ensure that all strands are firmly
clamped in the terminal block and that there is no excess bare
conductor showing, and no stray strands.
Summary of Contents for FlowCERT LITE
Page 1: ...FlowCERT LITE UL INSTRUCTION MANUAL...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...Page 6 EC DECLARATION OF CONFORMITY...
Page 18: ...Page 10 The full dimensions of the enclosure are as shown below...
Page 37: ...Page 29 This page left blank intentionally...
Page 77: ...Page 69...
Page 135: ...Page 127 This page left blank intentionally...