Chapter 2
Installation
Power Supply Requirements
FlowCERT
LITE
can operate from AC supply or from a DC battery and is
designed for use in temperatures between -4
o
F to +140
o
F (-20
o
C to +50
o
C).
The
AC
is
115V+5%/-10% 50/60Hz
. The
DC
is
22-28V
. In all cases the
FlowCERT
LITE
w
ill typically consume 6W of power, with a maximum of
10W. If
FlowCERT
LITE
has both an AC and DC supply available then the
AC supply source will be automatically sensed and used, should the AC
supply be removed for any reason then the DC supply will take over.
The AC and DC wiring should be completed using either 16 – 14 AWG
(1.5–2.5mm
2
) stranded or 16 – 14 AWG (1.5–4mm
2
) solid wire, with all
terminals being tightened to 4.5in lbs (0.5Nm).
An external supply isolator/circuit breaker (AC or DC) must be fitted near to
the unit and labelled to identify the instrument to which it refers.
Safety Symbols
Detailed below are descriptions and meanings of safety/warning symbols
that are used on the
FlowCERT
LITE
and in this manual.
Summary of Contents for FlowCERT LITE
Page 1: ...FlowCERT LITE UL INSTRUCTION MANUAL...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...Page 6 EC DECLARATION OF CONFORMITY...
Page 18: ...Page 10 The full dimensions of the enclosure are as shown below...
Page 37: ...Page 29 This page left blank intentionally...
Page 77: ...Page 69...
Page 135: ...Page 127 This page left blank intentionally...