Page 26
Using the RS232 Serial Interface
The RS232 serial interface is used to communicate between the
FlowCERT
LITE
and a PC using the optional Ultra PC and other associated Pulsar
software packages, to obtain information such as data logging and view echo
traces upload, download and save parameter files. In addition it can also be
used to control or obtain information using a standard PC or other computer
base equipment. To do so, the settings for control are as follows:
baud rate
19,200
,
8 data bits
,
no parity
,
1 stop bit
.
The device should be connected, via the serial port, as shown in
Chapter 2
Installation
.
To use the device remotely, you need to
log on
to start, and
log off
when
finished. When
logged on
,
FlowCERT
LITE
will show ‘
Remote ON
’ on
the display, and “
Communicator OFF
” when
logged off
.
All commands should be followed by a carriage return
.
The unit will respond either OK (or a value) if the command is accepted, or
NO if it is not.
To log on, send the command
/ACCESS:pppp where pppp is the passcode (P922).
To log off, send the command
/ACCESS:OFF
To read a parameter value, send the command
/Pxxx where xxx is the parameter you wish to read, and the
FlowCERT
LITE
will respond with the parameter value.
To set a parameter, send the command
/Pxxx:yy where xxx is the parameter number, and yy is the value you wish
to set it to.
Summary of Contents for FlowCERT LITE
Page 1: ...FlowCERT LITE UL INSTRUCTION MANUAL...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...Page 6 EC DECLARATION OF CONFORMITY...
Page 18: ...Page 10 The full dimensions of the enclosure are as shown below...
Page 37: ...Page 29 This page left blank intentionally...
Page 77: ...Page 69...
Page 135: ...Page 127 This page left blank intentionally...