P872 Fill Indicator
This parameter determines the rate at which the LCD
fill
indicator activates.
Default = 10m/min
P873 Empty Indicator
This parameter determines the rate at which the LCD
empty
indicator
activates.
Default = 10m/min
Rate
P874 Rate Update
This parameter determines the way in which the rate is calculated. If set to
continuous
(
P874
=
0
), then the rate is calculated and displayed
continuously, i.e. any change seen from shot to shot is calculated and
displayed, but if set to use
values
P874
=
1(Default)
then the
values
set in
P875
and
P876
are used to calculate and display the rate.
P875 Rate Time
This parameter is the period (in seconds) over which the material level rate
of change is averaged before the
Rate
Value
(
P877
) is updated. If the
Rate
Distance
(
P876
) is exceeded before the
Rate Time
(
P875
) has expired then
the
Rate Value
(
P877
) will be updated immediately.
Default = 60sec.
P876 Rate Distance
This parameter is the rate
Measurement Units
(
P104
) over which the
material level must change before the
Rate
Value
(
P877
) is updated. If the
Rate Time
(
P875
) expires before the
Rate Distance
(
P876
) is exceeded
then the
Rate Value
(
P877
) will be updated immediately.
Default = 0.05m
P877 Rate Value
This parameter displays the current rate of change of material level, in
Measurement Units
(
P104)
per minute. It is read only.
P878 Lower Cutoff
This parameter is used to select the minimum Rate to be calculated, and can
be used to eliminate unwanted updates from effects of ripples/waves on the
surface of the material.
Summary of Contents for FlowCERT LITE
Page 1: ...FlowCERT LITE UL INSTRUCTION MANUAL...
Page 2: ......
Page 4: ......
Page 8: ......
Page 14: ...Page 6 EC DECLARATION OF CONFORMITY...
Page 18: ...Page 10 The full dimensions of the enclosure are as shown below...
Page 37: ...Page 29 This page left blank intentionally...
Page 77: ...Page 69...
Page 135: ...Page 127 This page left blank intentionally...