2-6
VIBXPERT 11.2010
Temperature
This interface is used for the connection of a thermocouple (type K).
Communication
The green channel is used as the:
-
Communication interface to the PC. The connection is established
via a network (Ethernet) or directly via a USB cable.
-
USB printer port.
The network is configured in the 'Device Setup' (see 'Ethernet' on
page 2-34).
Interfaces on VIBXPERT with intrinsic safety
The interface functions are basically the same as in the non-EX
device. They differ in the following ways:
-
VIBXPERT with intrinsic safety has a separate charging socket on
the bottom of the device. Thus, the channels marked in blue are
used for measuring analog signals only.
-
The interface for the network and USB connection is located on
the bottom of the device and is not used in hazardous areas.
-
The communication adapter (VIB 5.330UNV) must be connected
between VIBXPERT and the peripheral device (PC, printer, net-
work) for communication (see page 2-32ff).
Getting started - Interfaces
PC
Printer
Network
PC
Communication
Digital
VIB 5.330-UNV
RS 232
USB - slave
USB - master
RJ 45
B
A
Channel A
Digital
Temperature
USB / Ethernet
Charging
socket
top view
bottom view
Channel B
Display