2-20
VIBXPERT 11.2010
Getting started - Basic settings
*'Show' appears for factory-defined
transducers whose parameters cannot
be changed.
1
only for VIBREX / VIBRONET
² only for 'Voltage' signal type
Transducer parameters
To check the parameters for transducers that have already been set
up, proceed as follows:
• Select the transducer.
• Press the MENU key.
• Click on 'Show'* or 'Edit' to open the parameter screen (see
below).
M
EAS
.
QUANTITY
Measurement quantity of the sensor: vibration acceleration, veloc-
ity, displacement, current, RPM, jump, user-defined quantity
S
IGNAL
TYPE
Signal type of the sensor (e.g. LineDrive, ICP, voltage)
I
NPUT
VOLTAGE
RANGE
±3V / ±8V
1
/ ±30V²; only applies for transducers of 'Voltage' and
'VIBREX/VIBRONET' signal types.
Q
UANTITY
(
USER
DEF
.)
Identification for a user-defined measurement variable.
U
NIT
(
USER
.
DEF
.)
Unit for the user-defined variable.
P
RECISION
Accuracy in decimal places.
C
OEFFICIENT
A
4,
A
3,
A
2
Parameters for linearizing a sensor's non-linear characteristic line
(e.g. MNS12 proximity probe - VIB 6.640).
S
ENSITIVITY
The sensitivity of the sensor determines the signal level.
O
FFSET
Offset of the sensor
L
INEAR
FROM
/
TO
The linearity range of the sensor and the filter setting of the
measurement must be adapted to one another.
Transducer parameters
for VIB 6.140 accelerometer