5-7
VIBXPERT 11.2010
Measurement tasks
F
A.1 Selecting the measurement task via the measurement task icon:
• Highlight the respective icon in the 'Advanced' tab.
• Press the F key to display the stored measurement tasks
• Select the required measurement task using the identifier.
The identifiers of the measurement tasks set up in the factory can
help in selection:
-
D
UAL
or 2-C: 2-channel measurements
-
S
PEC
or
SPECTRUM
: 'Spectrum RPM' measurement type
-
O
VERALL
V
ALUE
: 'Overall Value - RPM' measurement type
-
S
TARTUP
: Runup
-
C
OASTDOWN
: Coastdown
A.2 Selecting the measurement task in the Task Manager:
• Highlight the respective icon in the 'Advanced' tab.
• Press the MENU key and click on 'Task Manager'.
• Click in the uppermost menu to display the stored measurement
tasks. The 'Show Dual Tasks' or 'Show Single Tasks' menu entries
display the available 2-channel or 1-channel measurement tasks
(see below)
• Select the desired measurement task.
• Select the desired measurement task.
B. Starting the measurement
• Click on the measurement task icon.
A dialog window appears in the measurement screen in which you
can temporarily change the following setup parameters:
-
S
TART
RPM
, S
TOP
RPM
-
RPM
DEVIATION
-
M
EASUREMENT
RANGE
.
• Click on 'Start' after the parameters have been set correctly.
During data collection, the relevant measurement data are dis-
played in a dialog window (see below).
Left:
Runup / Coastdown
Phase - RPM (Bode plot)
Right:
Results screen during
a coastdown measurement
Phase - RPM