3-31
VIBXPERT 11.2010
Prewarning if measurement values deviate excessively
In the ‘Route’ mode, VIBXPERT recognizes by how much the current
measurement value differs from the last recorded historical measure-
ment value. If this deviation is large, the green LED on the device
lights up and the symbol R!
R!
R!
R!
R! appears in the results screen.
For this prewarning to appear, the following requirements must be
met:
-
The maximum permissible deviation of the measurement values is
set in the OMNITREND software (here: 20%).
-
The ‘% deviation’ and ‘Historical data’ functions are activated in
the OMNITREND Route Editor.
-
The information and data sets have been loaded into VIBXPERT
with the route.
-
The display mode for route results is set to ‘Standard’ (see also
Device settings for route, p. 2-15).
Example:
If the current result deviates from the
previous result by more than 20%, the
green LED lights up.
previous current
result
Trend