2-13
VIBXPERT 11.2010
Display settings, switchoff and additional options
• Click on 'Display' in the 'Device Setup' screen.
C
ONTRAST
AND
BRIGHTNESS
: Click on the respective field and move the
joystick to the right or left. To accept the setting, press the joystick.
S
WITCH
BACKLIGHT
OFF
/ A
UTO
POWER
OFF
: The display illumination / the
device automatically switches off if no action occurs within the
turn-off time (illumination: 10s ... 1m / device: 6m ... 6h). The
automatic switch off is disabled in the measurement screen /
results screen.
B
ACKGROUND
IMAGE
: A background image (PNG format) can be dis-
played in the start screen. Four standard images are stored on the
device. Additional images can be loaded onto the device using the
'VIBXPERT utility' tool (see also the appendix, p. 6-8).
S
HOW
BATTERY
CHARGE
: Display the remaining battery charge in percent.
S
PECTRUM
: S
HOW
LINES
BELOW
LOWER
FREQUENCY
(
F
MIN
): In spectra with a
lower limit frequency of f
min.
> 0 Hz, the lines between 0 Hz and
f
min.
are displayed in the diagram if this option is enabled.
H
IDE
CALIBRATION
HINT
: The message for overdue calibration is not
shown.
• To accept the settings, press MENU and click on 'OK'.
Protective display film
To protect the display against damage, it is covered with a
scratchproof transparent film. The film is easy to remove.
In the VIBXPERT with explosion protection, you must remove this
film before using the measurement device in explosive areas.
Device Setup
Getting started - Basic settings
Attention!
ZOOM
CLR
ENTE
R
GOTO
PAGE
Protective display film
Display settings
Summary of Contents for VIB 5.300 EX Series
Page 170: ...6 24 VIBXPERT 11 2010 Decleration of conformity ...