VIBXPERT 11.2010
Getting started - Basic settings
Service menu
• Click on the 'Service' icon to open the Service main menu (see
below).
A series of functions are available for service, maintenance and
training purposes. For clarity, the numerous settings are grouped into
submenus (see below):
Device Info
This menu contains device information for service and repair as well
as the date of the next offset compensation and calibration (for more
details, see also p. 6-16).
Offset Compensation
Various factors, e.g. aging or temperature, can cause an offset in the
analog electronics. This option makes it possible to maintain the
measurement accuracy of the instrument by regularly (approximately
every 2 months) compensating the offset.
• Click on 'Offset Compensation'.
• Click on ‘Start'. The process takes about 3 minutes.
Factory Defaults
This menu can be used to reset various settings in the instrument to
the values set up in the factory. In addition, data that are no longer
required in the instrument, e.g. language files, can be deleted.
• Click on 'Reset'.
!! RISK of DATA LOSS !!
Pay particular attention to which settings are reset or which data
are deleted. This action cannot be undone!
Device Setup
Device Setup
Device Setup
Device Setup
Attention!
Left:
Service main menu
Right:
Factory defaults