1049766 Rev A
V60 Ventilator Service Manual
10-1
Chapter 10.
Specifications
10.1 Control settings
Table 10-1 lists ventilator control setting ranges, resolutions, and accuracies.
Table 10-2 lists the controls active in the different ventilation modes.
Table 10-1: Control setting ranges, resolutions, and accuracies
Parameter
Range
Resolution
Accuracy
Factory default
Mode settings
Modes
CPAP, S/T, PCV,
AVAPS (optional)
N/A
N/A
S/T
Control settings
C-Flex
OFF, 1 to 3
1
N/A
2
CPAP
4 to 25 cmH
2
O
1 cmH
2
O
± (2 cmH
2
O + 4% of
target)
4 cmH
2
O
EPAP
4 to 25 cmH
2
O
1 cmH
2
O
± (2 cmH
2
O + 4% of
target)
4 cmH
2
O
IPAP
4 to 40 cmH
2
O
1 cmH
2
O
± (2 cmH
2
O + 4% of
target)
12 cmH
2
O
I-Time (Inspiratory
Time)
0.30 to 3.00 sec
0.05 sec
± 0.03 sec
1.00 sec
Max P (AVAPS Maxi-
mum IPAP Pressure)
6 to 40 cmH
2
O
1 cmH
2
O
± (2 cmH
2
O + 4% of
target)
25 cmH
2
O
Min P (AVAPS Mini-
mum IPAP Pressure)
5 to 30 cmH
2
O
1 cmH
2
O
± (2 cmH
2
O + 4% of
target)
10 cmH
2
O
O
2
(Oxygen)
21 to 100%
1%
± 5%
21%
Ramp Time
OFF, 5 to 45 min
5 min
± 1 sec
OFF
Rate (Respiratory Rate)
4 to 60 BPM
1 BPM
± 1 BPM
4 BPM
Rise (Rise Time)
1 to 5
1
N/A
3
V
T
(AVAPS Target Tidal
Volume)
200 to 2000 mL BTPS
5 mL
± 15%
500 mL