Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-7
8.2 Preliminary
Cleaning, Inspection
and Setup
WARNING:
To prevent disease transmission, use protective equipment when handling
contaminated bacteria filters or other patient accessories.
Before servicing the ventilator, clean and inspect as follows:
•
Clean the ventilator exterior as described in the
User Manual.
•
Remove the humidifier from the ventilator, if applicable.
•
Visually inspect the exterior of the ventilator for damage. Replace
damaged parts as needed.
•
Remove and inspect the cooling fan filter and air inlet filter. Clean or
replace filter as described in the
User Manual
.