Chapter 5
5-14
V60 Ventilator Service Manual
1049766 Rev A
Diagnostic Mode and Troubleshooting
5.2.4
Miscellaneous
The Miscellaneous screen allows you to access the significant event log, touch
screen diagnostics, download mode, high pressure leak test, and system leak
test.
1. From the Service screen, touch the
Misc
tab.
2. The screen includes the
Significant Event Log
,
Download Mode
,
Touch
Screen Diagnostics
,
High Pressure Leak Test
, and
System Leak Test
buttons (Figure 5-13).
Figure 5-13: Miscellaneous Screen
3. To exit the Miscellaneous screen at any time, touch
Back to
Diagnostics Menu
.