Chapter 6
6-20
V60 Ventilator Service Manual
1049766 Rev A
Reports and Software Downloads
6.3 Clearing the
Significant Event Log
The #CLRLOG command allows you to use your PC to clear the significant
event log.
CAUTION:
Only qualified service technicians are to clear the significant event log.
NOTE:
Do not clear the significant log until you have used the DRPTA command
to save the information to your PC using HyperTerminal or Tera Term.
1. Enter Diagnostic mode by pressing and holding the nav-ring
Enter
button and turn on the ventilator by pressing the
On/Shutdown
button.
Within 5 seconds, press the nav-ring
Enter
button again to enter
Diagnostic mode.
2. Connect the 25-pin to 9-pin adapter and null modem cable between
the ventilator COM port and the service PC.
3. Click on the HyperTerminal or Tera Term icon to launch the program
that communicates with the V60 ventilator.
4. Type:
#CLRLOG
(all caps) in the dialog box, and press
Enter
to clear
the diagnostic log (Figure 6-34).
Figure 6-34: Clear Diagnostic Codes Command
5. When the diagnostic log has been successfully cleared, the dialog box
displays this response:
?CLRLOGOK