1049766 Rev A
V60 Ventilator Service Manual
A-1
Appendix A Respi-Link
This section describes the following Respi-Link functions:
•
Downloading V60 ventilator software
•
Installing a V60 ventilator option
•
Remote troubleshooting
A.1 Downloading
Ventilator Software
Follow these steps to download V60 ventilator software using Respi-Link.
NOTE:
Do not connect the null modem cable until instructed to do so.
1. Turn on the PC and verify that the Gateway interface is running by
double-clicking on
Start Auto Respi-Link Gateway
in the Respi-Link
folder on your desktop.
2. If the ventilator is not already in Diagnostic mode, press and hold the
nav-ring
Enter
button and turn on the ventilator by pressing the
On/Shutdown
button on the user interface. Within 5 seconds, release
and press the nav-ring
Enter
button again to enter Diagnostic mode.
3. Connect a standard 9-pin male-female RS-232 null modem cable and
9-pin to 25-pin female-male adapter between the PC serial port and
the V60 ventilator.
4. Within 30 seconds, the GatewayInterface screen appears on the PC
monitor if a software package is pending (Figure A-1).