Chapter 8
8-22
V60 Ventilator Service Manual
1049766 Rev A
Performance Verification
5. Set the analyzer's pressure range to measure at least 60 cmH
2
O, and
connect the pneumatic calibration analyzer to the ventilator as shown
(Figure 8-15).
NOTE:
Zero the analyzer transducers, if required.
Figure 8-15: Pressure Accuracy Test Configuration
6. Configure the diagnostic screen (Figure 8-16) as follows:
•
Pressure:
0 cmH2O
•
Flow:
0 SLPM
•
O2:
21%
•
Solenoid 1:
None
•
Solenoid 2:
Machine Pressure Line
•
Solenoid 3:
Machine Pressure Line
•
Solenoid 4:
Prox
Pressure