Appendix A
A-8
V60 Ventilator Service Manual
1049766 Rev A
9. Touch
Service
and verify the Options on the V60 ventilator are the
same as the options listed on the GatewayInterface screen.
10. Click
Exit
on the GatewayInterface screen.
11. Turn the V60 ventilator off by pressing the
On/Shutdown
button.
12. Option installation is complete.
A.3 Reinitiate a
Software Package
Follow these steps to reinitiate a V60 ventilator software package using
Respi-Link.
NOTE:
Do not connect the null modem cable until instructed to do so.
1. Turn on the PC and verify the Gateway Interface is running by double-
clicking on
Start Auto Respi-Link Gateway
in the Respi-Link folder on
your desktop.
2. If the ventilator is not already in Diagnostic mode, press and hold the
nav-ring
Enter
button and turn on the ventilator by pressing the
On/
Shutdown
button on the user interface. Within 5 seconds, release and
press the nav-ring
Enter
button again to enter Diagnostic mode.
3. Connect a standard 9-pin male-female RS-232 null modem cable and
9-pin to 25-pin female-male adapter between the PC serial port and
the V60 ventilator.
4. Double click on the GatewayInterface icon to re-launch the install
screen (Figure A-11).
Figure A-11: GatewayInterface icon