Chapter 4
4-2
V60
Ventilator
Service Manual
1049766 Rev A
Periodic Maintenance
4.1 Annual
Preventive
Maintenance
Instructions
Follow these steps to perform the annual preventive maintenance (PM)
procedure on the V60 ventilator.
NOTE:
The annual PM procedure must be performed by a qualified service
technician in an appropriate setting, such as a biomedical repair facility.
Use a clean patient circuit and filters when performing any testing or
operational checkout.
1. Clean the exterior of the V60 ventilator using universal precautions
(including mask, gloves, eye protection).
2. Enter Diagnostic mode by pressing and holding the nav-ring
Enter
button and turn on the ventilator by pressing the
On/Shutdown
button
on the user interface. Within 5 seconds, press the nav-ring
Enter
button again to enter Diagnostic mode. The Diagnostics Menu appears
(Figure 4-1).
Figure 4-1: Diagnostics Menu