Chapter 7
7-24
V60 Ventilator Service Manual
1049766 Rev A
Component Removal/Installation
7.22 Oxygen Inlet
Filter
Follow these steps to remove the oxygen inlet filter (Figure 7-23). Reverse to
install.
NOTE:
When installing the oxygen inlet filter, ensure that the smaller end of the
filter is opposite the threads in the GDS.
1. Remove GDS (section 7.20).
2. Use a wrench to loosen the oxygen inlet filter and remove from GDS.
Figure 7-23: Removing the Oxygen Inlet Filter
Gas delivery subsystem (GDS)
Oxygen inlet filter
(362 + 34 N cm/
32 + 3 in.-lbf)