Chapter 8
8-26
V60 Ventilator Service Manual
1049766 Rev A
Performance Verification
14. Verify that the set air flow values correspond to the flow
measurements:
NOTE:
When using the Certifier FA Plus the air delivery/flow accuracy test
requires that the temperatures of the delivered gas and ambient air must
be within 10
o
C of each other, and that the relative humidity of the gas
must be less than 30% at 21.1
o
C.
Most air flow analyzer flow sensors are sensitive to relative humidity, and
typically indicate higher-than-actual air flow readings when the relative
humidity is above 10%. Consult the manufacturer’s instructions for
correcting any air flow inaccuracies due to high relative humidity.
15. Remove the low-pressure plug and verify that the flow reading on the
analyzer drops by more than 15 SLPM.
16. Touch
Pressure
then
Accept
to zero the flow setting.
17. The air delivery/flow accuracy test is complete.
Set
Flow
to:
Verify that analyzer and
average
Air Flow
displays read:
10 SLPM
9 to 11 SLPM
120 SLPM
108 to 132 SLPM
230 SLPM
207 to 253 SLPM (N/A if above 6400 feet)