Chapter 8
8-2
V60 Ventilator Service Manual
1049766 Rev A
Performance Verification
Table 8-2: Performance Verification Test Requirements
Service performed
Required tests
Annual preventive maintenance
All
AC inlet: removal/replacement
Electrical safety
Air and oxygen flow sensor assembly: removal/
replacement
Electrical safety
High pressure leak test
System leak test
Pressure accuracy
Air delivery/flow accuracy
Oxygen flow accuracy
Oxygen accuracy
Breath rate
Backlight inverter cable: removal/replacement
Electrical safety
Ventilator controls
Backlight inverter PCBA: removal/replacement
Electrical safety
Ventilator controls
Blower: removal/replacement
Electrical safety
System leak test
Pressure accuracy
Air delivery/flow accuracy
CPU PCBA: removal/replacement
Electrical safety
Alarms
Data acquisition (DA) PCBA: removal/
replacement
Electrical safety
High pressure leak test
System leak test
Pressure accuracy
Air delivery/flow accuracy
Oxygen flow accuracy
Oxygen accuracy
Breath rate
Data acquisition (DA) PCBA to flow sensors
cable: removal/replacement
Electrical safety
Data acquisition (DA) PCBA to motor controller
(MC) PCBA cable: removal/replacement
Electrical safety
Fan: removal/replacement
Electrical safety