Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-3
Gas delivery subsystem (GDS): removal/
replacement
Electrical safety
High pressure leak test
System leak test
Pressure accuracy
Air delivery/flow accuracy
Oxygen flow accuracy
Oxygen accuracy
Breath rate
Gas outlet port: removal/replacement
Electrical safety
System leak test
Pressure accuracy
Internal battery: removal/replacement
Electrical safety
Internal battery
LCD: removal/replacement
Electrical safety
Ventilator controls
LCD cable: removal/replacement
Electrical safety
Ventilator controls
Motor controller (MC) PCBA: removal/
replacement
Electrical safety
Air delivery/flow accuracy
Internal battery
Power fail
Nav-ring: removal/replacement
Electrical safety
Ventilator controls
Oxygen inlet filter: removal/replacement
Electrical safety
High pressure leak test
System leak test
Pressure accuracy
Oxygen flow accuracy
Oxygen accuracy
Oxygen valve: removal/replacement
Electrical safety
High pressure leak test
System leak test
Pressure accuracy
Oxygen flow accuracy
Oxygen accuracy
Breath rate
Power harness cable: removal/replacement
Electrical safety
Internal battery
Power management (PM) PCBA: removal/
replacement
Electrical safety
Internal battery
Power supply: removal/replacement
Electrical safety
Table 8-2: Performance Verification Test Requirements
Service performed
Required tests