Chapter 10
10-4
V60 Ventilator Service Manual
1049766 Rev A
Specifications
10.3 Alarms
Table 10-4 lists the adjustable alarm ranges and resolutions.
10.4 Menu window
settings
Table 10-4: Adjustable alarm ranges and resolutions
Parameter
Range
Resolution
Factory default
Hi Rate (High Rate
Alarm)
5 to 90 BPM
1 BPM
30 BPM
Lo Rate (Low Rate
Alarm)
1 to 89 BPM
1 BPM
10 BPM
Hi V
T
(High Tidal Vol-
ume Alarm)
200 to 2500 mL
BTPS
5 mL
2500 mL
Lo V
T
(Low Tidal Vol-
ume Alarm)
OFF, 5 to 1500 mL
BTPS
5 mL
OFF
HIP (High Inspiratory
Pressure Alarm)
5 to 50 cmH
2
O
1 cmH
2
O
50 cmH
2
O
LIP (Low Inspiratory
Pressure Alarm)
OFF, 1 to 40 cmH
2
O
1 cmH
2
O
OFF
Lo V
E
.
(Low Minute
Ventilation Alarm)
OFF, 1 to 99 L/min
BTPS
1 L/min
OFF
LIP T (Low Inspiratory
Pressure Delay Time
Alarm)
5 to 60 sec
1 sec
20 sec
Table 10-5: Menu window settings and ranges
Parameter
Range
Brightness
1 to 5
Loudness
1 to 10
Mask/ET Selection
ET/Trach, 1, 2, 3, 4, Other
Exhalation Port Selection
DEP (Respironics Disposable Exhalation Port), Whisper
Swivel (Respironics Whisper Swivel), PEV (Respironics
Plateau Exhalation Valve, Other (Other Exhalation Port),
None (No inline circuit exhalation port)
Screen Lock
Off, On