Chapter 8
8-46
V60 Ventilator Service Manual
1049766 Rev A
Performance Verification
Test 2: Leak Tests
Pass value
Fail value
Value after repair
High pressure leak test, oxygen inlet pressure at 45-50 psig
Analyzer pressure display (45-50 psig)
psig
psig
psig
Oxygen inlet pressure display (45-50 psig)
psig
psig
psig
Threshold
psig
psig
Pressure drop
psig
psig
psig
System leak test, machine pressure at 55-60 cmH
2
O
Threshold
cmH
2
O
Pressure drop
cmH
2
O
cmH
2
O
cmH
2
O
Test 3: Controls
Circle one
Circle one
AC power indicator
PASS
FAIL
Touch screen
PASS
FAIL
ON/Shutdown LED green
PASS
FAIL
Brightness control
PASS
FAIL
Nav-ring
PASS
FAIL
Screen lock
PASS
FAIL
Enter button
PASS
FAIL
ON/Shutdown LED amber
PASS
FAIL
Test 4: Pressure Accuracy
Pass value
Fail value
Value after
repair
Barometric Pressure
Analyzer barometric pressure display
mmHg
Barometric pressure
display mmHg
mmHg
mmHg
Pressure at 0 cmH
2
O
Analyzer pressure display (-1.5 to 1.5 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Avg
Machine
display (-1.5 to 1.5 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Avg
Proximal
display (-1.5 to 1.5 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Pressure at 1 cmH
2
O
Analyzer pressure display (0.5 - 1.5 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Avg Machine
display (0.5 - 1.5 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Avg
Proximal
display (0.5 - 1.5 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Pressure at 35 cmH
2
O
Analyzer
pressure
display
(32.5
-
38.5
cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Avg
Machine
display (32.5
-
38.5
cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Avg
Proximal
display (32.5
-
38.5
cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O