Chapter 7
7-4
V60 Ventilator Service Manual
1049766 Rev A
Component Removal/Installation
5. Holding the battery so that the vent hole faces up and the Philips logo
faces out, thread the battery cable through the battery bracket.
Position and place the battery inside the battery compartment.
Pinching the end of the battery connector, plug it in so that it locks in
place.
6. Reinstall the battery bracket by replacing the two screws. Reinstall the
side panel and secure the fastener with a 1/4 turn clockwise.
7. Make sure the battery is properly installed by plugging the ventilator
into an AC power receptacle and verifying that the yellow Battery
(charged) LED on the front panel flashes. The flashing LED indicates
the battery is being charged.
8. If necessary, attach the option label.
NOTE:
When installing a new battery:
•
The battery must be charged for at least 5 hours before being placed
into service.
•
Dispose of the old battery according to your institution’s protocol.
Follow all local, state, and federal regulations with respect to
environmental protection.
Figure 7-3: Removing the Internal Battery
M4 x 8 screw (x2)
(113 + 11.3 N cm/
10 + 1 in.-lbf)
Left side wall
Battery bracket
Internal battery
Install battery with vent hole on
top and Philips logo facing out