Chapter 5
Diagnostic Mode and Troubleshooting
1049766 Rev A
V60 Ventilator Service Manual
5-13
5.2.3
Output Controls
This Output Controls screen allows you to set and verify wrap, analog, and
digital outputs.
NOTE:
When entering wrap counts, ensure that the displayed values are within
+ 23 counts of the set value.
1. From the Service screen, touch the
Outputs
tab.
2. The screen shows diagnostic information and output controls (Figure
5-12).
Figure 5-12: Output Controls Screen