Appendix A
A-6
V60 Ventilator Service Manual
1049766 Rev A
16. Remove the null modem cable and adapter from the V60 ventilator.
17. Turn the V60 ventilator off by pressing and holding the
On/Shutdown
button on the user interface.
18. Enter Diagnostic mode by pressing and holding the nav-ring
Enter
button and turn on the ventilator by pressing the
On/Shutdown
button
on the user interface. Within 5 seconds, press the nav-ring
Enter
button again to enter Diagnostic mode.
19. Touch
Service
and verify that the Software Version is the same as the
version listed on the GatewayInterface screen (Figure A-8).
20. Click
Exit
on the GatewayInterface screen.
21. Turn the V60 ventilator off by pressing the
On/Shutdown
button, then
touch
Ventilator
Shutdown
.
22. Software download is complete.
A.2 Installing
Ventilator Options
Follow these steps to install V60 ventilator options using Respi-Link.
NOTE:
Do not connect the null modem cable until instructed to do so.
1. Turn on the PC and verify the Gateway Interface is running by double-
clicking on
Start Auto Respi-Link Gateway
in the Respi-Link folder on
your desktop.
2. If the ventilator is not already in Diagnostic mode, press and hold the
nav-ring
Enter
button and turn on the ventilator by pressing the
On/Shutdown
button on the user interface. Within 5 seconds, release
and press the nav-ring
Enter
button again to enter Diagnostic mode.
3. Connect a standard 9-pin male-female RS-232 null modem cable and
9-pin to 25-pin female-male adapter between the PC serial port and
the V60 ventilator.