Chapter 7
7-22
V60 Ventilator Service Manual
1049766 Rev A
Component Removal/Installation
7. Remove screw that attaches the ground cable to the manifold.
8. Remove the oxygen filter cap and air inlet collar from the right side
wall.
NOTE:
When reinstalling, apply a thin coat of Krytox to the inlet collar
flange on the side wall. Tighten the air inlet collar first, and then the
oxygen filter cap.
9. Disconnect the DA-MC cable from the DA PCBA.
10. Remove GDS from base.
NOTE:
When reinstalling, connect tubing so that:
•
Longer (9-in.) tube connects between manifold “M”
(machine) barb and barbed gas outlet fitting.
•
Shorter (8-in.) tube connects between manifold “P”
(proximal) barb and barbed proximal pressure port.
Figure 7-21: Removing the GDS
DA PCBA: remove
DA-MC cable (not shown)
Manifold cap
GDS manifold:
remove ground
cable (not shown)