Chapter 7
7-14
V60 Ventilator Service Manual
1049766 Rev A
Component Removal/Installation
7.13 Proximal
Pressure Port
Follow these steps to remove the proximal pressure port (Figure 7-13). Reverse
to install.
1. Remove front panel (section 7.10).
2. Remove the nut and lock washer that attach the proximal pressure
port to the inside of the front panel. (Proximal pressure port includes
the nut and lock washer.)
NOTE:
When reinstalling the proximal pressure port, install the ground wire
first, then the lock washer and nut.
Figure 7-13: Removing the Proximal Pressure Port
Nut (x1)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
Proximal
pressure port
(Inside front panel)
Ground wire
Lock washer (x1)