Chapter 6
Reports and Software Downloads
1049766 Rev A
V60 Ventilator Service Manual
6-23
5. The V60 ventilator stores the serial number and the dialog box
displays this response:
?CFGOK
6. Touch
Misc
then
Vent Info
, and verify that the serial number displayed
on the Vent Info screen matches the ventilator serial number.
7. Programming the serial number is complete.
6.5.2
Programming Ventilator Power-On Hours
NOTE:
Do
not
connect the null modem cable until instructed to do so.
Follow these steps to re-program the ventilator power-on hours to a new CPU
PCBA:
1. If the ventilator is not already in Diagnostic mode, press and hold the
nav-ring
Enter
button and turn on the ventilator by pressing the
On/Shutdown
button. Within 5 seconds, release and press the nav-ring
Enter
button again to enter Diagnostic mode.
2. Turn on the PC and then connect a standard 9-pin male-female
RS-232 null modem cable and 9-pin to 25-pin female-male adapter
between the PC serial port and the V60 ventilator.
3. Click on the HyperTerminal or Tera Term icon to enable
communication with the V60 ventilator.
4. Type:
#SETOPTIME
(all caps) with the additional information listed in the table below
separated by a comma (Table 6-1).
Table 6-1: Power-On Hours Command Format
Field
Description
Operating time ID
0 = Total power-on hours
1 = CPU PCBA
2 = PM PCBA
3 = MC PCBA
4 = DA PCBA
5 = Oxygen flow sensor PCBA
6 = Air flow sensor PCBA
Hours
Must be 6 digits, zero padded (for example:
001234), up to a maximum of 596523.