E-022-4-E
8
⑦
After filling in all the fields required, click on "OK" button. A message box as shown in Fig. 26 then
appears.
Clicking on "OK" at this window changes the previous settings to the new settings just entered.
However, the flowmeter output also changes with the changes in settings made. For safety's sake,
therefore, in applications where the flowmeter output controls a valve or other devices, it is
necessary that the control loop be switched to manual control to ensure that the control loop is
unaffected by the flowmeter output.
⑧
When the previous settings are replaced with the new settings just entered in response to clicking on
"OK"
,
a message box as shown in Fig. 27 appears. Clicking on "OK" button at this point completes
the setup.
Fig.26
Fig.27
⑨
After clicking on "OK" button, the message box for filling in fields of menu items appears again.
Click on "Cancel" button to hide the message box for filling in fields of setup items.
To abort the setup process, click on "Cancel" button in the course of steps
②
through
⑦
.
CAUTION
When the units of measurement are changed, it is necessary to change the units of
measurement included in other message boxes. When Some message box is open with
units of measurement included, close that message box once and then open it again
before renewing the units.