3.7.10 Density Calibration (Density cal)
Described in this section is the procedure to calibrate the density of the flowmeter transmitter.
①
Click on
"
Diag/Service (T)
"
at top
-
level menu of the screen, select
"
Density cal
"
at
"
Calibration
"
drop
-
down list, and click on again.
②
A message box like the one shown in Fig. 0 appears. To start the density calibration, fill the sensor
flow tube with density calibration fluid, key in the density value of calibration fluid, and click on
"
OK
"
button.
③
A message box as shown in Fig. 102 appears. If you want to make a density calibration, click on "OK"
.
To abort it, click on "cancel"
.
Fig.0
Fig.02
④
It takes about 30 seconds to complete the process of density calibration. While the density calibration
is in progress, a message like the one shown in Fig. 03 stays on.
Fig.03