E-022-4-E
46
⑤
When zeroing has come to an end, a message box like the one shown in Fig. 00 appears.
Click on "OK" button.
Fig.00
⑥
In response to clicking on "OK" button, the screen returns to the Auto Zero message box.
Click on "Cancel" button to hide the Auto Zero message box.
CAUTION:
It is necessary that the zero point adjustment be conducted while the process fluid is
stable and at the operating temperature. The process fluid inside the sensor unit must be
completely at zero flow. Otherwise, precise zeroing cannot be achieved.
For further details, refer to the information under the topic "Zero Point Adjustment" in the
high-accuracy OVAL Coriolis flowmeter instruction manual.