3.7.11 Trim Analog Output 1 (Trim Analog 1)
Analog output 1 of the flowmeter transmitter is trimmed at this menu. In the "Trim Analog Output 1"
,
adjustment is made by allowing outputs in analog values corresponding to the 4mA and 20mA, regardless
of the conditions of the process. For safety's sake, in applications where the flowmeter output controls a
valve (s) or other devices, it is necessary that the control loop be switched to manual control to ensure
that the control loop is unaffected by the flowmeter output.
①
Click on "Diag/Service (T)" at top-level menu of the screen, select "Trim Analog 1"
, and click on again.
②
A message box as shown in Fig. 106 appears. If you want to trim Analog Output 1, click on "OK"
button.
CAUTION:
Following this adjustment, the analog output from the transmitter is provided on the
basis of the reading of the test instrument (ammeter or voltmeter) connected. For this
reason, the test instrument to be coupled must be one that has been calibrated and is
accurate enough. The analog output is accurately adjusted in the factory before shipment
and requires no further adjustment.