E-022-4-E
42
Fig.87
⑤
In response to clicking on "Stop" button, a message box like the one shown in Fig. 87 appears.
Click on "OK" button.
⑥
Clicking on "OK" button returns the screen to the Fix Pulse message box. Click on "Cancel" button to
hide the Fix Pulse message box.
3.7.6 Loop Test of Status Output (Fix Status output)
At this menu, a loop test of the output line is conducted with the status output in the state of a simulated
output. Since a simulated output is used here, irrespective of the status of the process, in applications
where the flowmeter output controls valves or other devices, it is necessary for the sake of safety that the
control loop be changed to manual control.
①
Click on "Diag/Service (T)" at top-level menu of the screen, select "Fix Status output" from the "Loop
test" drop-down list, and click on again.
②
A message box like the one shown in Fig. 88 appears. Select either simulated output status "ON" or
"OFF"
,
and click on "Start" button.
Fig.88
③
A message box as shown in Fig. 89 appears. If you want to use a simulated output for the status
output, click on "OK."
To abort the process, click on "Cancel"
.
Fig.89