3.6.4 Analog Output 2
①
Click on "Setup (S)" at the top-level menu of the screen, select "Analog Output 2" from "Config output"
drop-down menu, and click on again.
②
Similar to the procedure outlined in Steps
②
through
⑤
in "3. Analog Output 1"
, Analog output 2 is set
up here.
3.6.5 Pulse Output
①
Click on "Setup (S)" at the top-level menu of the screen, select "Pulse output" from "Config output"
drop-down menu, and click on again.
②
A message box like the one shown in Fig. 3 appears. Pulse output 2 is set up at this window.
③
You are to set up individual items.
The low cutoff is "OFF" when your "Lowcut" option is "0.0%"
.
Default setting is "0.0%"
.
If "Bi direction" is chosen, the cutoff functions both in the forward and reverse direction.
④
After filling in all the fields required, click on "OK" button. A message box as shown in Fig. 32 then
appears.
Clicking on "OK" at this window changes the previous settings to the new settings just entered.
However, the flowmeter output also changes with the changes in settings made. For safety's sake,
therefore, in applications where the flowmeter output controls a valve or other devices, it is necessary
that the control loop be switched to manual control to ensure that the control loop is unaffected by the
flowmeter output.
⑤
When the previous settings are replaced with the new settings just entered in response to clicking on
"OK"
,
a message box as shown in Fig. 33 appears. Clicking on "OK" button at this point completes the
setup.
⑥
After clicking on "OK" button, the message box for filling in fields of items appears again. Click on
"Cancel" button to hide the message box for filling in fields of items. To abort the setup process, click
on "Cancel" button in the course of steps
②
through
④
.