3.7.3 Loop Test of Analog Output 1 (Fix Analog 1)
At this menu, a loop test of the output line is conducted with the analog output in the state of a simulated
output. Since a simulated output is used here irrespective of the status of the process, in applications
where the flowmeter output controls valves or other devices, it is necessary for the sake of safety that the
control loop be changed to manual control.
①
Click on
"
Diag/Service (T)
"
at top
-
level menu of the screen, select
"
Fix Analog
"
from the
"
Loop test
"
drop
-
down list, and click on again.
Click here.
Fig.79
Fig.80
②
A message box like the one shown in Fig. 80 appears. Select the simulated output level and click on
"Start" button. If you want to obtain any current output in analog value (4 to 20mA) other than "4mA"
and "20mA"
,
select "Other" key in the analog value for the output desired, and click on "Start"
.
Fig.8
③
A message box as shown in Fig. 8 appears. If you want to use Analog Output for the simulated
output, click on "OK"
.
To abort the process, click on "Cancel"
.
To obtain a simulate output
other than 4mA and 20mA,
click here and enter
appropriate value.