After filling in all the fields required, click on "OK" button. A message box as shown in Fig. 59 then
appears. Clicking on "OK" at this point changes the previous settings to the new settings just entered.
However, the flowmeter output also changes with the changes in settings made. For safety's sake,
therefore, in applications where the flowmeter output controls a valve or other devices, it is necessary
that the control loop be switched to manual control to ensure that the control loop is unaffected from
the flowmeter output.
⑤
When the previous settings are replaced with the new settings just entered in response to clicking on
"OK"
,
a message box as shown in Fig.60 appears. Clicking on "OK" button at this point completes the
setup.60
⑥
Following clicking on "OK" button, a message box for entering cert. items appears again. Click on
"Cancel" button to hide the message box for entering setup items.
To abort the setup, click on "Cancel" button in the course of steps
②
through
④
.
3.6.13 Flow Calibration Factor (Flow Cal)
①
Click on "Setup (S)" at top-level menu of the screen, select "Flow Cal" from "Charize sensor" drop-
down list, and click on again.
②
A message box as shown in Fig. 6 appears. At this window, the flow calibration factor is set up.
CAUTION:
Bore size has been set up according to the sensor used and no further change in bore
size is required.