Fig.6
Fig.66
③
After filling in all the fields required, click on "OK" button. A message box as shown in Fig. 65 then
appears.
Clicking on "OK" at this window changes the previous settings to the new settings just entered.
However, the flowmeter output also changes with the changes in settings made. For safety's sake,
therefore, in applications where the flowmeter output controls a valve or other devices, it is necessary
that the control loop be switched to manual control to ensure that the control loop is unaffected by the
flowmeter output.
④
When the previous settings are replaced with the new settings just entered in response to clicking on
"OK"
,
a message box as shown in Fig. 66 appears. Clicking on "OK" button at this point completes the
setup.
⑤
After clicking on "OK" button, the message box for filling in fields of individual items appears again.
Click on "Cancel" button to hide the message box for filling in fields of items.
To abort the setup process, click on "Cancel" button in the course of steps
②
through
③
.
CAUTION:
Density factors have already been set up before shipment from the factory and no further
adjustment is required. Do not attempt to change the settings unless recalculating is
absolutely necessary.
In the EV9201, menu item "Density Cal" is not included.
3.6.15 Zero Factor
Zero factor is a value which is written down during the zeroing; no further adjustment is required in normal
use. Only in the event sensor diameter is changed, it is required to set the value to 0 and perform the
zeroing. Item "Xmtr auto zero" is always set in the "OFF".
CAUTION:
A zero factor modified to an improper value will result in inaccurate flow measurement.
Do not attempt to change the setting except when sensor size is changed.