Troubleshooting
Section 9-3
Common Check Items for Error Codes E160 to E182
• Check that the orientation of the code is within
10
.
• Check that the code size is set correctly.
• Check that the model has been registered correctly using the
Reference Model procedure given on page 65.
9-3
Troubleshooting
Connection Errors
The power indicator is not lit.
• The power supply is not connected properly.
• The supply voltage is not 24 VDC+10%/–15%.
The RUN indicator is not lit.
• The cable is not connected correctly. (Use Image Analysis/
Hardware Check to confirm.)
The OK indicator is not lit.
• The cable is not connected correctly. (Use Image Analysis/
Hardware Check to confirm.)
• Teaching has not been performed.
• Use a monitor to be sure the code is picked up by the Scanner.
E180
E181
E182
Decode error
An L-shape finder pattern was detected but the
data could not be decoded.
• The data part of the code may be stained or
damaged. Check the condition of the data
part of the code.
• The finder pattern may be stained or dam-
aged. Check the condition of the finder pat-
tern.
• Model registration may not have been per-
formed in the optimum way. Re-register the
model with reference to page 65.
• The cell recognition coordinates may be
incorrect. Adjust the field of vision so that
the code image is larger and re-enter the
settings.
E200
Timeout error
Decoding was not possible within the timeout
time.
• Redo teaching or re-register the pattern to
obtain suitable setting.
• Increase the timeout time.
Error Code
Description
Countermeasures