3. Operation Mode and Status
3-4
3.3. Specifications of Operation Mode and Status
Operation Mode/Status
Definition
Transition to Other Status
Startup Process
The system is under startup
processing in normal mode or
safe mode after turning on the
power or reset (pressing the
button or Web UI operation).
When an instrument failure occurs,
the status transitions to "system
error status". For other cases, see
below:
- If the setup DIP switch
SW10=OFF, the status transitions
to one right before turning on the
power or reset.
- If the setup DIP switch
SW10=ON, the status transitions
to "safe mode".
Normal
Mode
Setup Status
A status in which setup by
EQ-Manager is available.
If the system is started up under
normal mode, the status
becomes the setup status.
- The status transitions to
"communication testing status" by
Web UI operation or by the
operation of EQ-Manager.
- The status transitions to
"preparing for collecting" then
"collecting status" by pressing the
RUN/STOP button(*), Web UI
screen, or EQ-Manager
operations.
- When an instrument failure or a
setup/status failure occurs, the
status transitions to "system error
status".
Communication
Test
Only data collecting from
measurement devices is
continuously performed.
- The status transitions to "setup
status" by Web UI or EQ-Manager
operation.
- When an instrument failure or a
setup/status failure occurs, the
status transitions to "system error
status".
Collecting status A status in which measured data
are collected from a
measurement device and
logged (in the internal memory).
Collecting is continued even
after an occurrence of a
communications error or
monitoring process error.
- The status transitions to "setup
status" by pressing and holding
the RUN/STOP button(*), or Web
UI or EQ-Manager operation.
- When an instrument failure or a
setup/status failure occurs, the
status transitions to "system error
status".
System Error
Status
A status in which an instrument
failure of the EQ100 occurred.
No transition occurs to other
statuses.
Safe Mode
A status in which the system
cannot be properly started up
due to a hardware failure or a
firmware update failure. Or a
status transitioned from other
status after turning on the power
with the setup DIP switch
SW10=ON or reset (pressing
the button or Web UI operation).
In this mode, operations of
setup initialization and firmware
update is available from the
Web UI screen.
It does not transition to normal
mode.
*: If the setup DIP switch SW9=OFF, the RUN/STOP button can be operated. If SW9=ON, the
RUN/STOP button is disabled.