part
3
83
Command Formats
cont.
COMM.PARAMETERS.210
= 000
300
BAUD
COMM. PARAMETERS.43
= 00
No Parity
001
600
01
Odd
010
1200
10
Even
011
2400
11 N/A
100
4800
101
9600
110 19200
COMM. PARAMETERS.5
= 0
7 Bit Data
COMM. PARAMETERS.6
= 0
1 Stop Bit
1
8 Bit Data
1
2 Stop Bit
Example:
Set Baud 9600, Odd, 7 Bit, 1 Stop. The command data is 00001101Bin = 0DHex. Send *W100D.
BUSFORMAT.0
= N/A
BUS FORMAT.1
= 0 No LF
BUS FORMAT.2
= 0 No Echo
1 LF
1 Echo
BUSFORMAT.3
= 0 RS-232
BUS FORMAT.4
= 0 Continuous
BUS FORMAT.5
= 0 CR
1 RS-485
1 Command
1 Space
Example:
Set Space, Continuous, RS-232, Echo, Line Feed, N/A. The command data is 00000110Bin = 06Hex. Send *W1F06.