2.5. Electrical Installation
POWER CONNECTION EXAMPLES
The figure below shows the power wiring hookup.
Fuse
Connector
Output Type
115 Vac
230 Vac
Fuse 1
Output 1
Relay
5 A(T)
3 A(T)
Fuse 2
Output 1
Relay
5 A(T)
3 A(T)
SSR
0.5 A(T)
0.5 A(T)
Fuse P
Power
N/A
100 mA(T)
100 mA(T)
Fuse A
Alarm 1/2
Relay
3 A(T)
3 A(T)
Note:
Values shown are minimum recommendations for the
protection of the controller. For a specific load, consult the respected
electrical specifications to select a suitable fuse.
The Protective Conductor terminal must be
connected for safety reasons.
The Safety European Standard EN61010-1 for measurement, control, and laboratory
equipment requires that fuses must be specified based on IEC127. This standard
specifies for a Time-lag fuse, the letter code ÒTÓ. The above recommended fuses are
of the type IEC127-2-sheet III. Be aware that there are significant differences
between the requirements listed in the UL 248-14/CSA 248.14 and the IEC 127 fuse
standards. As a result, no single fuse can carry all approval listings. A 1.0 Amp IEC
fuse is approximately equivalent to a 1.4 Amp UL/CSA fuse. It is advised to consult
the manufacturerÕs data sheets for a cross-reference.
Use copper conductors only for power
connections
Figure 2.5.1 Power Wiring Hookup
Note
☞
Summary of Contents for MICROMEGA CN77000 series
Page 4: ...NOTES ...
Page 19: ...SETUP part 2 11 Dip Switch Settings ...
Page 31: ...Figure 3 2 Flowchart for Input Type OPERATION CONFIGURATION MODE part 3 23 ...
Page 36: ...Figure 3 3 Flowchart for Reading Configuration part 3 OPERATION CONFIGURATION MODE 28 ...
Page 45: ...Figure 3 5 Flowchart for Loop Break OPERATION CONFIGURATION MODE part 3 37 ...
Page 49: ...OPERATION CONFIGURATION MODE part 3 41 Figure 3 6 Flowchart for Output 1 ...
Page 59: ...OPERATION CONFIGURATION MODE part 3 51 Figure 3 7 Flowchart for Output 2 ...
Page 65: ...OPERATION CONFIGURATION MODE part 3 57 Figure 3 8 Flowchart for Ramp Soak ...