7. PE1000Win software
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
86
No.600-2111-09
GLP
The GLP tab is shown below.
Lamp Active Time:
Time for which the mercury lamp has been on
Exhaust Filter Hg:
Amount of mercury absorbed on activated carbon for
absorbing measured mercury
Septum Count:
Counter for with the needle penetrates though the
septum.
Sample Injection Count: Number of times that the syringe has been operated
The figures in parentheses are threshold-warning values, with a warning being
issued when they are exceeded. The Clear button appears when a warning is
generated. Take the appropriate measures such as replacement, and then
press the button to clear it.
Operation Count:
Time for which the power has been on
Measurement Count:
Number of times of measurement
The above two items cannot be cleared.