7. PE1000Win software
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
61
No.600-2111-09
ANALYSIS screen
①
①
①
①
Profile window
Displays all measurement result profiles. The number of profiles in one window and the
height of the scale specified here are also used in the print profiles.
The scrollbar on the left of the window is used to switch to full scale.
The scrollbar on the bottom of the window is used to change the displayed data number.
The number shown at the bottom left of the window is the number of data displayed in
one window, which can be set to any number.
②
②
②
②
Calibration curve window
Displays the current calibration curve.
A prepared calibration curve is displayed on the “CALIB” tab.
Verification results of the calibration curve are shown on the “RESULT” tab.
The verification results of the calibration curve are most effective when a cubic formula
is selected, and displays the concentration or mercury weight results obtained from the
absorbance on which the calibration curve preparation is based, using a linear
calibration curve.
③
③
③
③
Table window
Shows the same information as that displayed in the Table window of the CONTROL
screen.