5. Measurement
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
30
No.600-2111-09
Setting the method
This section describes the settings of the various operating conditions of the sample
changer and the sample heating conditions.
There are some preset conditions, but they can be freely changed.
①
Click the cell of "Method Number" for opening the
list box.
②
Select a method from the list. (24items :0~23)
* Detailed settings can be made by selecting [Run]-
[Methods set] from the menu.
(Refer to 7.5.1Method setting)
Selecting Measure Range
Set the heating speed of the mercury collection section for measurement in accordance
with the weight of the mercury.
LOW: 0 to 10 ng (Low concentration) HIGH: 0 to 100 ng
Click the cell of "Meas. Range" for opening the
list box. And select the “Range”.
When not selecting a range, it measures in the range set up on "measurement
Conditions" (7.5.2Measurement condition (Mode)).
Setting the number of injections
Specifies the number of times the samples are to be injected.
Click the cell of "Multiple Injection”, and input a
count of injection.